Chapter 6: Analyzer Operation
To set check standard measurements (Scheduled)
1. On the
Configure Online
screen (
Setup
tab>
Online
button>
Configure Online
button), press the
Check
Standard
button. The
Check Standard
screen appears.
Figure 47: Check Standard Setup Screen
2. Press the
Frequency
button to set how often to measure the check standard.
3. Press the
Start Time
button to set the start time using a 24-hr. clock format (e.g., use 13:00 for 1 p.m.).
4. Press the
Concentration
button to set the ppm concentration of the check standard.
5. Press the
Tolerance
button to set the acceptable range of concentrations for the measurement of the
standard. For example, if the
Tolerance
is set at 20%, then the acceptable concentration is 20% below
or 20% above the
Concentration
value.
6. Press the
Check Std. Setup
button. The following buttons appear on the
Check Standard
screen.
Figure 48: Check Standard (with Setup Buttons) Screen