GE Analytical Instruments ©2016
21 of 270
DLM 68100-09 EN Rev. A
Warning
Hazardous reagents (sodium persulfate, hydrochloric acid, and phosphoric acid)
are used in the Analyzer. The waste stream from the instrument is acidic and must
be disposed of properly. Consult your federal, state, and local government
regulations.
Warning
This is a Safety Class I product. It must be must be attached to a grounded power
source. To avoid damage to the Analyzer and the chance of injury, make sure the
Analyzer is connected to a power line of the proper voltage.
Warning
To ensure proper operation and protect against injury, do not operate the Analyzer
with the covers off or the door open.
Warning
Any operation requiring access to the inside of the Analyzer, including installation of
maintenance items, could result in injury. To avoid potentially dangerous electric
shock or burning, turn off power and disconnect from the power supply before
opening the Analyzer. Use caution when touching any electronic components.
Warning
If a mounting stand accessory is used with the Analyzer, the mounting stand must
be bolted either to a wall or the floor prior to installation of the Analyzer.
Warning
The pressure of the compressed gas at the Analyzer's gas inlet must not exceed
100 psig (689 kPa). The pressure of compressed gas at the blowback gas inlet on
the Filter Panel accessory must not exceed 40 psig (276 kPa). Use only dry, oil-free
air or nitrogen.