Chapter 6: Analyzer Operation
Online Measurement—Configuration
Figure 45: Configure Online Screen
You can use the
Configure Online
screen (
Setup tab
Online button
Configure Online
) to set up online
measurements for the following:
• Operation with Multiple Streams
• Check Standards (Scheduled or Manual)
• Continuous (online measuring)
• Cycle Frequency
• Clean Inlet
• Auto Restart
• Daily Flush
• Daily Flush Start
You can also calculate Run Time and view a Summary of online configuration on this screen.
Operation with Multiple Streams
The InnovOx On-Line TOC Analyzer may be used to monitor multiple streams. The base configuration can
monitor two streams, and five streams can be monitored with the optional 5-Stream configuration. With the
exception of the number of streams, both Analyzers are set up the same way.
N
OTE
:
The
Fail Safe
button, previously located on this screen, has been moved to the
Advanced screen (Maintenance tab >
Advanced
). For information on this feature, refer