GE Analytical Instruments ©2016
166 of 270
DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
To set a location for the Analyzer
1. Press the
Location
button, and use the keyboard that appears to specify the location of the Analyzer.
2. Press
Set Location
. The
Advanced
screen appears again. Press the
Back
button to return to the
Maintenance
tab.
Setting System Properties
Calibration constants and other key parameters are stored in the Analyzer’s memory. However, you may need to
refer to the original factory settings. Before you begin customizing the parameters and running analysis data,
export and save these system properties.
To save and export the system properties
1. Select the
Maintenance
tab.
2. Press the
Advanced
button. The
Advanced
screen appears.
3. Press the
System Properties
button. The
System Properties
screen appears.
Figure 67: System Properties Screen
4. To overwrite the system properties with factory settings that you have previously exported to a USB
flash drive, insert the drive into the Analyzer and press the
USB Restore
button.
5. Press the
USB Backup
button or
Ethernet Export
button.