Parameter specifications
SpO
2
test methods used to establish accuracy claims during
low perfusion
Portrait
TM
technology with Portrait sensors P-SA01, P-SP01, P-W01, and P-SE01 has
been validated for SpO
2
low perfusion accuracy in bench top testing using a Fluke
Biomedical ProSim
TM
8 Vital Signs Simulator with signal strength setting of 0.2%
modulation and oxygen saturation levels of 70 to 100% at a pulse rate of 90 bpm.
SpO
2
test methods used to establish pulse rate accuracy
Portrait
TM
technology with Portrait sensors P-SA01, P-SP01, P-W01, and P-SE01 has
been validated for pulse rate accuracy in bench top testing using a Fluke Biomedical
ProSim
TM
8 Vital Signs Simulator at 0.2% modulation in the pulse rate range of 30 to
250 beats per minute (bpm) at SpO
2
of 90%.
SpO
2
performance specifications
The following specifications apply to all SpO
2
sensors.
SpO
2
measurement value and display
range
0 to 100%
SpO
2
measurement value accuracy
The accuracy is specified in terms of the root-mean-square (RMS)
difference between the measured values and the reference values
(A
rms
). Because pulse oximetry equipment measurements are
statistically distributed, only about two-thirds of the pulse oximetry
equipment measurements can be expected to fall within the ±1
A
rms
of the value measured by a CO-oximeter. For more detailed
information, refer to the SpO
2
supplemental analysis graphs.
●
SpO
2
(70 to 100%): ≤2%
●
Low perfusion SpO
2
(70 to 100%): ≤3%
●
With motion: SpO
2
(70 to 100%): ≤3%
SpO
2
display resolution
1 digit (% of SpO
2
)
Wavelength of SpO
2
sensor LEDs
This information can be especially useful to clinicians such as
those performing photodynamic therapy.
Pulse oximetry sensor LED peak wavelengths are within 600 to
1000 nm and the maximum optical output power for each LED is
less than 15 mW.
Pulse oximetry peripheral pulse rate
measurement (PR) display range
30 to 300 beats per minute
Pulse oximetry peripheral pulse rate
accuracy (A
rms
)
●
≤2 bmp (30 to 250 bpm)
●
Low perfusion: ≤2 bmp (30 to 250 bpm)
●
With motion: ≤5 bmp (30 to 250 bpm)
Pulse oximetry peripheral pulse rate
display resolution
1 bpm
Pulse oximetry SpO
2
and pulse rate
measurement data update period
< 30 s
158
Portrait
™
Mobile Monitoring Solution
5815148-01
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...