F
Electromagnetic compatibility
IEC 60601-1-2
The system complies with IEC 60601–1–2:2014.
According to parameter-specific standard requirements for ESU (electrosurgical unit)
tests, the equipment is protected against malfunction caused by electrosurgery.
EMC warnings
WARNING
DEGRADED PERFORMANCE. Do not use portable RF
communications equipment (including peripherals such as
antenna cables and external antennas) closer than 30 cm (12
inches) to any part of this device, including cables specified
by the manufacturer. Otherwise the performance of this
device/system may degrade.
WARNING
DEGRADED PERFORMANCE. Do not use an RFID reader
operating at a low frequency such as 134.2 kHz and 13.56
MHz close to any part of the Hub and sensors. Otherwise
there is a risk of degraded performance of the impedance
respiration measurement.
●
Impedance respiration measurement value may be out
of the measurement accuracy range and respiration
waveform noise may be randomly high when the
respiration sensor is too close to such RFID readers. When
these RFID emissions are removed, the measurement
returns to expected measurement accuracy within 60
seconds without any changes to settings.
WARNING
ELECTROMAGNETIC INTERFERENCE. Magnetic and electrical
fields are capable of interfering with the proper performance
of the device. For this reason make sure that all external
devices operated in the vicinity of the device comply with
the relevant EMC requirements. X-ray equipment or MRI
devices are a possible source of interference as they may
emit higher levels of electromagnetic radiation. Changes or
modifications to this device/system not expressly approved
by GE may cause EMC issues with this or other equipment.
This device/system is designed and tested to comply with
applicable standards and regulations regarding EMC and
needs to be installed and put into service according to the
EMC information stated as follows: This device/system is
suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes. Mains power should be that of a typical commercial
or hospital environment. Device is compliant to Class A.
5815148-01
Portrait
™
Mobile Monitoring Solution
189
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...