Battery management
1. MBAN connection
2. Sensor connection
3. Battery charge
The following table describes their functionality.
LED
Description and functionality
MBAN connection indicator.
●
Blinking amber slowly: the sensor battery is connected to the sensor,
but sensor has not been paired with the Hub.
●
Blinking amber quickly: MBAN connection to the Hub is lost.
●
Three slow blinks in green: the sensor battery is connected to the
sensor and the sensor has been paired with the Hub.
●
Blinking slowly in green, combined with an intensity change
sequence of all three LED indicators (brighter/dimmer): connection
check between the Hub and a sensor is ongoing.
●
Blinking sequence in light turquoise, combined with a blinking
sequence of all three LED indicators: boot-up sequence.
Sensor connection indicator.
●
Blinking amber slowly: the sensor battery is disconnected from the
sensor.
●
Blinking amber quickly: the connected sensor is faulty.
●
Three slow blinks in green: the sensor battery is connected to the
sensor.
●
Blinking slowly in green, combined with an intensity change
sequence of all three LED indicators (brighter/dimmer): connection
check between the Hub and a sensor is ongoing.
●
Blinking sequence in light turquoise, combined with a blinking
sequence of all three LED indicators: boot-up sequence.
Battery charge indicator.
●
Blinking amber with medium rate: the sensor battery has charge left
for less than 1 hour, but for more than 5 minutes.
●
Blinking amber quickly: the sensor battery has charge left for less
than 5 minutes.
●
Blinking slowly in green, combined with an intensity change
sequence of all three LED indicators (brighter/dimmer): connection
check is ongoing.
●
Blinking sequence in light turquoise, combined with a blinking
sequence of all three LED indicators: boot-up sequence.
74
Portrait
™
Mobile Monitoring Solution
5815148-01
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...