Cleaning, disinfection, and care
Post-cleaning inspection
Inspect devices visually after cleaning and prior to disinfection for the following:
●
Cleanliness
■
If there is any visible soil present, repeat the cleaning procedure until no visible
soil remains.
●
Damage, including, but not limited to, corrosion (rust, pitting), discoloration,
excessive scratches, flaking, cracks, and wear
●
Missing or worn device labeling or marking
Do not use devices that are not functioning properly or are damaged, or have
unrecognizable markings, or missing or worn device labeling or marking.
Cleaning procedure for sensor batteries
Initiate cleaning and disinfection as soon as possible after use (recommended within 1
hour). Remove any soil as soon as possible by wiping the device to prevent the drying
of soil on the device as described in following procedures.
Soiled devices must be separated from non-contaminated devices to avoid
contamination of personnel or surroundings.
Preparing a sensor battery for cleaning
1.
Remove the sensor and attachment accessories from the battery, or remove
the battery from the charger.
SpO
2
sensor
Respiration sensor
2.
Visually ensure that the device is intact. Evidence of damage may include, but is
not limited to, discoloration, excessive scratches, wear, and cracks.
Cleaning sensor batteries
In case there is some dried or gross soil, perform pre-cleaning by wiping with a soft,
clean non-linting cloth saturated with lukewarm tap water (temperature range 27 to
44°C, 81 to 111°F) until visible dried or gross soil is removed.
1.
Dampen a soft non-linting cloth with detergent or use ready-to-use detergent
wipes.
Prepare the detergent according to the manufacturer’s instructions.
122
Portrait
™
Mobile Monitoring Solution
5815148-01
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...