18
Alarm messages
Alarms related to the respiration rate
measurement
Alarm priorities are given in the section discussing alarm specifications.
Make sure that you are familiar with the generic layout of the Hub and Central Viewer
screens. This will help you identify where the following alarm messages appear. The alarm
location is indicated with the following abbreviations:
●
al. area = alarm area
●
data = data area
Physiological messages
Alarm message
Location
Possible explanations
Suggested actions
Hub:
●
Apnea
●
APN
Central Viewer:
●
Apnea
●
APN
●
al. area
●
data
No breathing detected during
the time that has been set as the
apnea limit.
●
Check the patient.
●
Check electrode placement
and skin to electrode contact.
Hub:
●
Resp artifact
●
Artifact
Central Viewer:
●
Resp artifact
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
Respiration signal quality is very
low due to patient movement or
poor electrode contact.
●
Check the patient.
●
Check electrode placement
and skin to electrode contact.
5815148-01
Portrait
™
Mobile Monitoring Solution
137
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...