Alarms
Safety precautions
Familiarize yourself with the safety chapter of this manual for all safety precautions
related to the system. Some warnings or cautions may also be listed with detailed
instructions in this chapter.
Alarm conditions
There are two types of alarm conditions: physiological and technical.
Physiological alarm conditions are triggered by a patient measurement being outside
the parameter limits or by apnea. Only high and medium priority alarms are available
for clinical review in trends data or in the Single Patient View events. You can find
more information on them in the alarm log files.
Technical alarm conditions are triggered by an electrical, mechanical, or other failure
of the equipment, or by failure of a sensor or component.
Alarm priority levels
Priority level
Color
Priority definition
High
Red
Requires an immediate response
Medium
Yellow
Requires a prompt response
Low
Cyan (blue)
Requires awareness of this condition
Informational
Gray
Provides additional information and is not an actual
alarm.
Active alarms
Active alarms for a patient are sorted from top to bottom, by acknowledgement and
then by priority and then by time. Unacknowledged alarms are sorted to the top
of the list in a priority order. Within each priority group the most recent alarms are
sorted higher.
On the Hub only one alarm is visible at a time in the alarm area. To see more alarms,
select
.
5815148-01
Portrait
™
Mobile Monitoring Solution
83
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...