Alarm messages
Alarm message
Location
Possible explanations
Suggested actions
Hub:
●
Learning
●
data
The patient’s breathing pattern
is being learned because the
measurement has just been
started or because of very large
artifacts in the signal.
●
Replace the respiration patch.
●
Wait until the message
disappears.
Hub:
●
RR high
●
RR low
Central Viewer:
●
RR high
●
RR low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient.
●
Check electrode placement
and skin to electrode contact.
●
Adjust alarm limits on the
Central Viewer if necessary.
Technical messages
Alarm message
Location
Possible explanations
Suggested actions
Hub:
●
Check Resp patch
●
Check patch
Central Viewer:
●
Check Resp patch
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
The respiration patch is not
properly connected to the
patient or the sensor is not
properly connected to the
respiration patch.
●
Check the respiration patch
placement and skin to
electrode contact.
●
Check the sensor and
respiration patch connection.
Hub:
●
Resp battery low
Central Viewer:
●
Resp battery low
●
al. area
Sensor battery charge is low.
●
Charge the battery.
●
Replace the battery.
Hub:
●
Change Resp battery
Central Viewer:
●
Change Resp battery
●
al. area
Sensor battery is empty.
●
Replace the battery and
put the empty one into the
charger.
Hub:
●
Clean Resp battery
interface
●
Clean battery
interface
Central Viewer:
●
Clean Resp sensor
interface
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
There is dirt in the sensor or
battery.
●
Check and clean the sensor
and battery.
Hub:
●
No Resp battery
●
No battery
Central Viewer:
●
No Resp battery
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
Battery has been removed from
the sensor.
●
Connect a sensor battery.
138
Portrait
™
Mobile Monitoring Solution
5815148-01
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...