Troubleshooting
Problem
Solution
●
No connection to patient monitoring services
appears: the Hub is not connected to the Core
Services. Make sure that the Hub is within the
network coverage range. If that does not help,
contact service personnel.
●
No license available
appears: there are no
available monitoring licenses. Unassign one Hub
or wait until a license becomes available again.
Monitoring is lost
Environmental interference or configuration errors
prevent the MBAN from working properly.
●
Check the environment for any interfering devices
and remove as necessary.
●
If the problem persists, contact qualified service
personnel to check the MBAN configuration.
There may be problems with power supply (cables,
batteries etc.) or with the WLAN.
●
Check battery charges and replace batteries as
needed.
●
Contact qualified service personnel.
Loss of audible and/or visual alarms
●
Check that the alarm limits are set correctly for
your care area.
●
Check the alarm volume adjustment.
●
Check the use mode (
REMOTE
or
LOCAL
) and
change as needed.
●
Check patient connections.
●
Contact qualified service personnel.
Waveforms seem incorrect or have artifact
There are several possible reasons for this and the
solution depends on the cause:
●
Patient movement: Instruct the patient about the
effect of movement on the measurement.
●
Environmental interference: Check the
environment for any interfering devices and
remove as necessary.
●
Sensor out of range: Reposition the Hub closer to
the sensor.
●
Sensor blocked: Reposition the sensor.
●
Broken or damaged sensor: Check the sensor and
replace as necessary.
●
Poor lead or sensor preparation, placement,
or attachment, or dry electrode gel: Check
sensor position and connections and correct as
necessary.
Incorrect or missing trends
Sensor is out of range.
●
Reposition the Hub closer to the sensor.
5815148-01
Portrait
™
Mobile Monitoring Solution
133
Summary of Contents for Portrait Hub01
Page 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Page 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Page 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Page 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Page 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Page 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Page 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Page 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Page 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Page 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Page 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Page 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Page 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Page 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Page 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Page 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Page 197: ...content...