82
8.2.2 Development unit
Dimensions
300 x 300 x 138 mm (W x L x H)
Weight
4.8 kg
Capacity
1 or 2 gels
Agitation
Gels are rotated in solutions during development
Development
Solutions are automatically pumped into and out of the
development chamber
Number of ports
9 ports are available for solutions to enter and exit the
chamber through; port 0 i reserved for waste.
Temperature control
range
Up to 50°C; the chamber only heats.
time
_
<4 min. to heat solutions from 20 to 45°C.
error
±4°C from programmed temperature once the programmed
temperature is reached.
Development chamber
material
Stainless steel
volume
Approximately 70 ml of solution will be pumped into the
chamber.
Temperature sensor
Stainless steel.
Level sensor
Enclosed in glass.
Gel holder
Stainless steel.
8.2.3 Common data
Safety regulations
EMC
This product meets the requirements of the EMC Directive
89/336/EEC through the harmonized standards EN 50082-
2 (emission) end EN 50082-1 (immunity).
Note: The declaration of conformity is valid for the
instruments when it is
• used in laboratory locations
• used in same state as it was delivered from
GE Healthcare Bio-Sciences AB
except for alterations
described in the user manual.
• used as ”stand alone” unit or connected to other CE
labelled
GE Healthcare Bio-Sciences AB
instruments or
other products as recommend.
Safety regulation
This product meets the requirements of the low Voltage
Directive (LVD) 73/23/EEC through the harmonized
standard EN 61010-1.
Safety glass
Class 1 apparatus
Operation environment
room temp.
4-40°C
humidity
Max. 95%.
Electricity requirements
mains voltage
100/120 V AC (120 model)
220-230/240 V AC (220 V model)
mains frequency
50-60 Hz
power
Maximum;
30 VA Separation and control unit
200 VA Development unit
330 VA together
Power disturbance
The system is protected against mains disturbance and
static discharges.
Mains failure
For port failures lasting less than 5-10 seconds, running
methods will automatically continue when power i returned.
For power failures lasting more than 5-10 seconds, running
methods are set to pause, an alarm sounds, and a message
appears on the display informing you about the power
failure.
8. Ordering information and technical data