GE H
EALTHCARE
D
IRECTION
2286865, R
EVISION
14
LOGIQ™ 7 S
ERVICE
M
ANUAL
5-42
Section 5-7 - Air Flow Control
Section 5-7
Air Flow Control
5-7-1
Air Flow Distribution
The four air flow pathes allow the scanner to be cooled down as shown in the figure above.
•
Path A (Front lower left > Filter > PC Box > Rear upper left) for PC Box cooling.
•
Path B (Front lower left > Filter > LV unit > Rear lower left) for LV unit cooling.
•
Path C (Bottom right > Filter > NEST Assy > Rear upper right) for NEST Assy cooling.
•
Path D (Rear bottom right > Filter > HV unit > Rear upper right) for HV unit cooling.
5-7-2
Filters
The scanner contains the three filters located at:
-
Front lower left for air flow of the LV unit and PC box.
-
Bottom right for air flow of the NEST Assy.
-
Rear bottom right for air flow of the HV unit.
Figure 5-67 Air Flow Inside the Scanner
A
B
C
D
A
B
C
D
Summary of Contents for LOGIQ 7
Page 2: ......
Page 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Page 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Page 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Page 393: ......