GE H
EALTHCARE
D
IRECTION
2286865, R
EVISION
14
LOGIQ™ 7 S
ERVICE
M
ANUAL
4-6
Section 4-2 - General Procedure
4-2-2-3-2
R 4.x.x to R6.0.x
1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON.
2.) The “Probing for maintenance access” dialog appears. Enter
Password
, then click on
OK.
3.) Then Start Loader display will be shown on the Monitor display.Click on
Maintenance.
4.) Click on
Exit to Windows
in maintenace diaglog window.Then the scanner is booted up
automatically.
4-2-2-3-3
R2.x.x to R4.x.x
1.) Insert the service dongle in the Service port located at the rear panel.
2.) The Start Up Screen will be shown on the Monitor display when the system is turned ON.
3.) Then Start Loader display will be shown on the Monitor display. Then the scanner is booted up
automatically.
NOTE:
To enter the Maintenance Mode, select Maintenance button.
NOTE:
Start is selected automatically when it time out.
Figure 4-8 Clicking on Mainttenance
Figure 4-9 Start Application Window