Section 6-7 - Cleaning the Trackball
Section 6-7 Cleaning the Trackball
(cont’d)
5.) Wipe off any oil or dust from the trackball using a cleaner or dry cloth.
6.) Wipe off any oil or dust from the trackball housing, rollers, and small ball, using a cleaner or cotton
bud.
NOTICE
When cleaning the trackball housing, make sure not to spill or spray any liquid into the trackball housing
(Keyboard or system).
Avoid organic solvents that may damage the mechanical parts of the trackball assembly.
Do not apply much force to the small ball.
Figure 6-43 Cleaning TrackBall and Housing
Trackbal Housing
Cotton Bud
Roller
Roller
Small Ball