Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10-4
Check LCD Monitor Hardware Setting
For 19 inch Monitor
To adjust the brightness:
1.) Press the Toggle button (2) or (3)
Once
for brightness adjustment.
a.) Verify that Brightness is
80 (Dark Room) or 100 (Bright Room)
.
If it is not, increase brightness to press the adjustment button (3).
If it is not, decrease brightness to press the adjustment button (2).
b.) Press the button (1) Once. The brightness display disappears.
NOTE:
19 inch LCD is ONLY a brightness adjustment.
8-2-10-5
Software Option Check
1.) Touch
Utility > Admin.
2.) Click on
System Admin
tab and verify that
HW Number
is the same as the Option Dongle ID
.
Figure 8-97 19 inch LCD Monitor Adjustment buttons
Figure 8-98 HW Number
1