After application is successfully installed, various LOGIQ™ 7 system parameters must be properly set.
Table 8-7 System Settings
Item
Method
Proper Value
Software Version
Utility
\ System
\ About
s/w version : R8.x.x
Image Part No. :
5305206-3
Regional Preset
Utility
\ System
\ About
\ Additional About Information
For the system other than “NONE”
selected as Regional Preset, resetting
MUST be required.
Computer Name
Utility
\ Connectivity
\ TCP/IP
\ Computer Name
\ Enable DHCP
\ IP Address
\ Subnet Mask
\ Default Gateway
Same as Serial Number
OFF
restore customer setting
restore customer setting
restore customer setting
Patient
Database Check
Utility
\ Connectivity
\ DataFlow
\ Default Data Flow
Press Patient (Keyboard or Touch Panel)
Local Archive – Int. HD
(n/a – Automatic)
LCD Monitor Setting
Press ‘Toggle Button’ for few seconds
(various)
Software Option
Utility
\ Admin
\ System Admin
\ HW Number
Same as Option Dongle ID
VCR
Utility
\ Peripherals
\ Video Setting
\ VCR
PAL/NTSC-depends on region
(set accordingly)
Language pre-work (For
Japanese Only)
Utility
\ Regional Option
\ Language (tab)
\ Install Files to East Asian Language
(Reboot)
ON / Apply