Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-29
8-2-9
Installing R8.x.x Application Software
8-2-9-1
Installing R8.x.x Application Software
1.) Power ON the scanner. The windows desktop appears.
2.) Insert the Application Software CD-ROM into the DVD drive.
NOTICE
You must install BT09 Base image before Application Software installation. If not, Application shows the
error message, “Service platform is out of date. You must reghost the C drive and reinstall the software.”
NOTICE
If you are attempting to install only application software, you MUST first check current Regional Preset
Value. Press
Utility > System > About
(Details in 4-2-6 - Archiving and Loading Presets for BT07
Figure 8-87 Desktop
NOTICE
At this time, the DVD writing software screen appears, but you can ignore the screen.
NOTICE
Wait approx. 20 seconds until the disk can be read.