LOGIQ™ 7 S
ERVICE
M
ANUAL
Section 10-3 - Periodic Maintenance Schedule
10-3
10-3-1 How often should PMs be performed?
Table 10-2 Periodic Maintenance Schedule
Service at Indicated Time
Daily
Weekly
Monthly
Per
Facilities
QA
Program
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
•
Clean Air Filter
•
more frequently depending on
your environment
Inspect AC Mains Cable
•
Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Clean Monitor and Touch Panel
•
Inspect Wheels, Casters, brakes and
Swivel Locks
•
Mobile Unit Check Daily
Check Control Panel Movement
•
Mobile Unit Check Daily
Console Leakage Current Checks
•
also after corrective
maintenance or as required by
your facilities QA program
Peripheral Leakage Current Checks
•
also after corrective
maintenance or as required by
your facilities QA program
Surface Probe Leakage Current Checks
•
also after corrective
maintenance or as required by
your facilities QA program
Endocavity Probe Leakage
Current Checks
also after corrective
maintenance or as required by
your facilities QA program
Transesphongeal Probe Leakage
Current Checks
also after corrective
maintenance or as required by
your facilities QA program
Surgical Probe Leakage
Current Checks
also after corrective
maintenance or as required by
your facilities QA program
Measurement Accuracy Checks
•
also after corrective
maintenance or as required by
your facilities QA program
Probe/Phantom Checks
•
also after corrective
maintenance or as required by
your facilities QA program