Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-39
8-2-10-6
VCR Parameters Setting (for the VCR-equipped system ONLY)
1.) Touch
Utility > System.
2.) Click on
Peripherals
tab.
3.) Set the proper parameters:
-
PAL or NTSC
-
VCR type
4.) Shut down the scanner, then restart the scanner.
5.) Verify that the VCR can be properly operated with no error (remote, play...) .
8-2-10-7
Regional Setup (For the system with language other than English ONLY)
For Japanese Languages
You can set up the keyboard for Japanese. If you select Japanese, the following items display or enter
by Japanese.
•
Patient name
•
Exam comment in Patient screen and Worksheet
•
Report/Direct Report free text, Insert Text
•
Annotation, Annotation Library
•
Error/Warning message
•
DICOM/Worklist, Query/Retrieve, Image Storage, Printer, MPPS
Figure 8-99 Peripherals