Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-15
4.) The following screen appears. Press
B
then
Enter
to initialize the
C
drive
5.) Check if the message appears as reloading C: partition only, then press
Enter
key.
If the message appears as reloading the entire disk, press
CTRL-ALT-DEL
to reboot the sysytem
and retry from procedures described in previous page.
Figure 8-65 Select B
CAUTION
If you select [A], all the archive would be lost.
Figure 8-66 Enter