Chapter 3 - Installation
Page 3-9
3-5-3
Installation of Two Probe Adapter & LOGIQ
α
200 Probe Adapter (Optional)
1.)
Remove the adhesive tape to open the flaps of two probe port or LOGIQ
α
200 Probe Adapter
box
2.)
Remove the PU Foams inside the two probe port or LOGIQ
α
200 Probe Adapter box (Refer
3.)
Take out the two probe port or LOGIQ
α
200 Probe Adapter (Refer Figure 3-6)
Note:
The two probe port or LOGIQ
α
200 Probe Adapter can be connected to or disconnected
from the system at any time regardless of whether the system is powered ON or OFF.
Figure 3-6 Unpacking Two Probe Adapter or LOGIQ
α
200 Probe Adapter
PU Foam
Box
Two Probe port
or
LOGIQ a 200 Probe Adapter