
G
E
HE
ALTHCARE
D
IRECTION
5305880-100, R
EVISION
3
LOGIQ™ 100 PRO S
ERVICE
M
ANUAL
Chapter 10 - Quality Assurance
Page 10-11
10-5-6-3
Basic Probe Cleaning
Refer to the User’s Manual for details on probe cleaning.
To help protect yourself from blood borne diseases, wear approved disposable gloves. These are
made of nitrile derived from vegetable starch to prevent allergic latex reactions.
Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s warranty.
DO NOT soak or wipe the lens with any product not listed in the User Manual. Doing so could result
in irreparable damage to the probe. Follow care instructions that came with the probe.Disinfect a
defective probe before you return it. Be sure to tag the probe as being disinfected.
Section 10-6 Electrical Safety Tests
10-6-1
Safety Test Overview
The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC
60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety
evaluation of cord-connected, electrically operated, patient care equipment. If additional information
is needed, refer to the NFPA 99 (For USA) and IEC 60601-1 documents.
THE USER MUST ENSURE THAT THE SAFETY INSPECTIONS ARE PERFORMED AT
LEAST EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF THE PATIENT
SAFETY STANDARD IEC-EN 60601-1. ONLY TRAINED PERSONS ARE ALLOWED TO
PERFORM THE SAFETY INSPECTIONS MENTIONED ABOVE.
To avoid electrical shock, the unit under test must not be connected to other electrical
equipment. Remove all interconnecting cables and wires. The unit under test must not be
contacted by users or patients while performing these tests.
Possible risk of infection. Do not handle soiled or contaminated probes and other components
that have been in patient contact. Follow appropriate cleaning and disinfecting procedures
before handling the equipment.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 182 of 197