3.5.1.1
System Software Installation through USB Thumbdrive:
1.) Switch OFF the LOGIQ™ 100 PRO unit.Connect the USB thumb drive to the USB port
provided in the rear panel.
2.) Switch ON the unit and wait till the system boots up completely and the B mode image is
seen.
3.) Now Switch OFF the LOGIQ™ 100 PRO unit and switch ON again.
4.) The system should display the latest version of software that was loaded through the USB.
The USB thumb drive would have the latest software in it,which would be sent by the Manufacturer
before software upgrade.
Whenever Upgrade of LOGIQ™ 100 PRO is done at field, Please ensure patient data back up
is done before upgrading the machine.
Do not remove the USB thumb drive when the image is getting transfered.
3-5-2
Installation of Optional Accessories
1.)
Unpack the foot switch and connect it to the connector on the rear panel. (Refer Figure 3-5)
2.)
Connect the VCR to a suitable power outlet. Using the BNC to BNC cable connect the Video
Out on the rear panel of LOGIQ™ 100 PRO to the Video IN of the VCR.
3.)
Connect the Video Graphic Printer (VGP) to a suitable power outlet. Using the BNC to BNC
cable connect the Video Out on the rear panel of LOGIQ™ 100 PRO to the Video IN of the
VGP. Connect the shutter cable from the VGP printer to the rear panel of the Logiq 100 PRO.
4.)
The images and report pages of the Logiq 100 PRO can be printed on HP Laser jet
printer(Model HP1020) through the USB port provided on the rear panel.Images can be
transfered to a PC using USB Thumb drive.
Note:
The LOGIQ™ 100 PRO has only one Video Out connection Therefore either a VGP
or a VCR only can be connected at a time
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 52 of 197