GE H
EALTHCARE
D
IRECTION
5305880-100, R
EVISION
3
LOGIQ™ 100 PRO S
ERVICE
M
ANUAL
Page 8-18
Section 8-5 - PCBs
8-5-2
Disassembly/Assembly of FEB for LOGIQ™ 100 PRO
Follow steps and refer
1.)
Remove Top Cover. Refer 8-2. For FRU Details refer 9-1-10
2.)
Remove the screw fixing the FEB PCB to the Rear panel bracket.
3.)
Remove four screws (a,b,c,d) from the probe connector.
4.)
Remove one screw which is fixing FEB to the chassis in the front.
5.)
Disconnect the two flat cables interfacing Davinci and Keyboard.
6.)
Remove all cables connected to the FEB and Davinci board on the rear of the system.
7.)
Hold the FEB and pull upward.
8.)
Remove four screws fixed on the Davinci board to mate the Davinci and the FEB.
9.)
Remove the Davinci board from the FEB Flush type connector slowly.
10.) For assembly follow the reverse order.
Table 8-8 Functional tests
Service Manual
Section
Functional Test/ Diagnostic Test
Debrief Script
4-2-1
Power ON/ Boot up
Service manual direction 5305880-
100, Section 8-5-2.Equipment
passed all required tests and is
ready for use.
4-2-1
Power OFF/ Shut down
4-3
Functional Checks
4-5
Software configuration checks
7-2-2
Diagnostics tests
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approved Document - 5305880-100TPH_r3.pdf Page 147 of 197