GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
8-100
Section 8-6 - Lower Section Components Replacement
Note:
below shows a close-up view of the up/down/swivel handle,
gas spring cable, and swivel release cable fitted in position, as seen from underneath .
7.) Using a combination wrench, release the swivel release cable from the cable retaining bracket by
loosening the securing nut, as shown in
8.) Release the swivel release cable from the up/down/swivel handle by carefully sliding the cable end
stopper out through the grooved slot on the handle assembly (refer to the detail shown in
9.) Repeat
to release the gas spring cable from the cable retaining bracket.
10.)Repeat
to release the gas spring cable from the up/down/swivel handle (refer to the detail
Figure 8-86 Handle and Gas Spring Cables - Close-up View
Figure 8-87 Releasing the Gas Spring Cable from the Cable Retaining Bracket
Cable retaining bracket
Handle
Swivel release cable
Cable securing nut
Gas spring cable
Swivel release cable