Section 7-3 - Performing Front End (FE) Diagnostics
5) Review the test results in the
Comment
area and proceed as follows:
•
If no problems were identified, proceed to re-boot the system, as prompted (
•
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in the
section, on page 6-11, and then repeat the RFI test.
•
If errors were identified and the results were not marginal, verify that the RFI cables are
connected and functioning properly, as described in
Chapter 5 - Components and Function
.
If the problem persists, contact your local GE Medical Systems field engineer for
assistance.
Figure 7-6 RFI Tab - Self Test Complete