GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-65
8-5-10
Keyboard Control Board Replacement Procedure
8-5-10-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the keyboard
control board replacement procedure.
8-5-10-2
Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in
.
8-5-10-3
Keyboard Control Board Removal Procedure
1) Remove the Vivid™ 4 left and right side covers, and rear cover - see the
on page 8 - 7 and the
Side Covers Installation Procedure
2) Disconnect the power cable from the rear of the BEP.
3) Remove the BEP side covers, as described in the
BEP2 - Side Cover Removal Procedure
4) If applicable, remove the octopus card holder as described in the
5) Disconnect the two keyboard control cables that are connected at the rear of the BEP, and all the
internal cables connected to the keyboard control board.
Note:
Counting from the top, the keyboard control board is positioned in the sixth PCI slot.
6) Release the securing screw located on the left of the board bracket.
7) Gently remove the keyboard control board from the BEP.
8-5-10-4
Keyboard Control Board Installation Procedure
1) Gently insert the keyboard control board into its slot.
Note:
Counting from the top, the keyboard control board is positioned in the sixth PCI slot.
2) Fasten the board with the screw on the left of the bracket.
3) Reconnect the two keyboard control cables that are connected at the rear of the BEP, and connect
the other internal cables to the keyboard control board.
4.) If applicable, install the octopus card holder as described in the
5) Install the BEP side covers, as described in the
BEP2 - Side Cover Installation Procedure
6) Reconnect the power cable to the rear of the BEP.
7) Return the Vivid™ 4 rear, left and right covers - see the
Side Covers Installation Procedure
Rear Cover Replacement Procedures
Note:
Perform a complete diagnostic of the system, as described in
.