Chapter 8 - Replacement Procedures
8-73
8-5-14
PC-VIC Replacement Procedure
NOTE:
This procedure is performed for systems with a BEP2 (Pentium 4-based BEP, newer type).
8-5-14-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the PC-VIC
replacement procedure.
8-5-14-2
Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in
.
8-5-14-3
PC-VIC Removal Procedure
1) Remove the left cover, as described
2) Remove the BEP side cover as described in the
BEP2 - Side Cover Removal Procedure
3.) Remove the screw that holds the Ground cable and PC-VIC in position (as shown in
4) Carefully slide the PC-VIC out from the unit.
5.) Disconnect all the cables at the rear of the PC-VIC module and the hard disk.
6.) Remove the four screws (two on each side) from the side of the PC-VIC module holder as shown
in
Figure 8-56 PC-VIC