Chapter 3 - Installation
3-49
3-6-10-2
Technical Support Tab - RFT Configuration
Monitoring
- should
always
be enabled. (this provides constant monitoring of critical
parameters to ensure safe operation of the system).
Watchdog
- normally should not be enabled. When enabled, if there is a system lock-up
caused by the Back End, this will turn OFF power at the Front End.
MOD Fan
- should be enabled only on BEPs based on Pentium IV - this provides faster
operation of the MOD. Enabling this option in the wrong type of BEP may
cause damage to the MOD or MO media.
Use Front End
- normally should be enabled.
ILinq
- should be enabled only on systems that use the iLinq option.
Figure 3-30 Technical Support Tab - RFT Configuration