Chapter 5 - Components and Function (Theory)
5-29
Section 5-5
Back End Processor
5-5-1
Introduction
The Back End Processor (BEP) supports the operation of the
Vivid™ 4 ultrasound unit and is the main
controller for the unit. The BEP comprises the various components described in the following sections:
•
•
•
Frame Grabber (RFI systems only)
•
•
Plug and Scan Card and Battery
•
•
•
•
•
•
•
•
•
on page 5-30 shows the location of the various components within the BEP for systems with
the RFI configuration. For systems with the RFT configuration, see
A block diagram of the BEP in Vivid™ 4 systems with the RFI configuration is provided in
on page 5-32. For systems with the RFT configuration, see