Patient Data
2000966-386D
Dash™ 3000/4000/5000
6-3
Up to 36 arrhythmia events and 10 ST events (including alarms and references) are
stored. Up to 15 events are displayed in the
ALARM HISTORY
information window
at one time. When monitoring from telemetry, the event is stored at the tower and the
last 36 events are viewable from the monitor.
The
ALARM HISTORY
contains 10 seconds of the arrhythmia event in a compressed
format. Due to the display size, only eight seconds of waveform data is shown. For
asystole, the first eight seconds of an event is displayed. All other arrhythmia events
display the last eight seconds of the event.
WARNING
MEASURING DATA STORED IN ALARM HISTORY—
Waveform data stored in
ALARM HISTORY
uses compression
technology that may not allow perfect reconstruction of the
waveform data when subsequently viewed. Although differences
occur relatively frequently, they are usually very minor. Users are
urged to verify diagnostic waveform measurements with the
waveform data from real-time graph strips.
Three leads of ECG are displayed for arrhythmia events. If an ART line was
monitored at the time of the event, two leads of ECG and the ART waveform are
displayed with the associated numerics.
To select an alarm or event, position the pointer in front of the option and press the
Trim Knob
control. Selecting an arrhythmia event displays the applicable waveform
and new menu options.
When viewing an ST alarm or reference, a QRS complex of all available leads
(except AVR) is displayed. If viewing an ST alarm, an asterisk is displayed next to
each lead in violation of ST limits.
Use the following options to review the
ALARM HISTORY
.
Vital Signs
VITAL SIGNS
displays the patient’s monitored parameter values for the last 24
hours, including the last 20 cardiac calculations and the last 10 pulmonary
calculations.
NOTE
If alarms were paused or the monitor was discharged, the word
OFF
displays in
the
ALARMS
field of the
VITAL SIGNS
information window during the time
either of these events occurred.
To review vital signs, select
MORE MENUS
>
PATIENT DATA
>
VITAL SIGNS
.
Option
Function
VIEW OLDER ALARM
View the previous event.
VIEW NEWER ALARM
View the next event.
DELETE ALARM
Delete the selected event.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......