Monitoring ECG
2000966-386D
Dash™ 3000/4000/5000
8-13
Arrhythmia alarm levels
HR
,
PVC
and
ST
parameter alarm levels
ECG SIZE
DETECT PACE
ARRHYTHMIA
ST ANALYSIS
LEAD ANALYSIS
HR HIGH LIMIT
and
HR LOW LIMIT
PVC LIMIT
and PVC state
To select the source of ECG settings, follow this procedure.
1.
Select
MORE MENUS
>
MONITOR SETUP
>
MONITOR DEFAULTS
>
SETUP DEFAULT DISPLAY
.
2.
Locate and select
COMBO DEFAULT SOURCE
.
3.
Select one of the following:
TELE DEFINED
— Use the telemetry ECG settings. This is the factory
default.
USER DEFINED
— Use the monitor’s ECG settings if telemetry is NOT
admitted. Use the telemetry ECG settings if telemetry IS admitted.
ECG Setting Source When Entering COMBO Mode
The following tables indicate the source (monitor or telemetry) of ECG settings when
entering
COMBO
mode.
Notice that selecting
Tele Defined
means the ECG setting source is always telemetry
when entering
COMBO
mode, while the ECG setting source for
User Defined
depends on whether the patient is admitted to the monitor or telemetry when entering
COMBO
mode.
Tele Defined
Monitor
Telemetry
Entering
COMBO
Mode
ECG Setting Source
Not admitted
Not admitted
>
Telemetry
Admitted
Not admitted
>
Not admitted
Admitted
>
User Defined
Monitor
Telemetry
Entering
COMBO
Mode
ECG Setting Source
Not admitted
Not admitted
>
Monitor
Admitted
Not admitted
>
Not admitted
Admitted
>
Telemetry
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......