10-6
Dash™ 3000/4000/5000
2000966-386D
Monitoring NBP
CAUTION
—The pulse rate derived from an NBP determination (measurement)
may differ from the heart rate derived from an ECG waveform
because the NBP parameter measures actual peripheral pulses, not
electrical signals or contraction from the heart. Differences may
occur because electrical signals at the heart occasionally fail to
produce a peripheral pulse or the patient may have poor peripheral
perfusion. Also, if a patient’s beat-to-beat pulse amplitude varies
significantly (for example, because of pulsus alternans, atrial
fibrillation, or the use of a rapid-cycling artificial ventilator), blood
pressure and pulse rate readings can be erratic, and an alternate
measuring method should be used for confirmation.
CAUTION
—For SuperSTAT NBP (Adult/Pediatric) Only — Simultaneously
monitoring ECG will enhance SuperSTAT NBP performance in
irregular rhythm.
NOTE
A patient’s vital signs may vary dramatically during the use of cardiovascular
agents such as those that raise or lower blood pressure or those that increase or
decrease heart rate.
Monitoring
To verify that you are ready to monitor NBP, follow this procedure.
1.
Admit the patient to the monitor.
2.
Connect a patient cable to the
NBP
patient connector on the monitor.
3.
Select a cuff appropriate for the limb size.
4.
Position the cuff on the patient.
5.
Verify the tubes between the cuff and the monitor are not kinked or blocked.
6.
Verify the correct cuff size is selected in the
NBP
menu.
7.
Start an NBP reading.
8.
Verify that patient data displays in the
NBP
parameter window.
9.
Adjust the control settings as necessary.
Control Settings
To adjust the control settings for the patient admitted to the monitor, select the
NBP
parameter window label. Then select any of the following options and make the
desired changes.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......