![GE Dash 3000 Operator'S Manual Download Page 52](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578052.webp)
3-4
Dash™ 3000/4000/5000
2000966-386D
Monitor Setup
COMBO
— Acquire ECG data from a monitor or a telemetry receiver cabinet
and access all the available parameters from the monitor. A network connection
is required.
ROVER COMBO
— Combines the mobility of
ROVER
mode and the telemetry
compatibility of
COMBO
mode.
To determine the mode running on a monitor, select
MORE MENUS
>
MONITOR
SETUP
>
SOFTWARE CONFIGURATION
. The monitoring mode is listed under
ADMIT MENU TYPE
.
Patient-Monitor Type
There are three patient-monitor types. The patient-monitor type controls the monitor
defaults and monitor settings. It is defined during installation via the password
protected
SERVICE MODE
menu.
ADULT-ICU
NEONATAL-ICU
OPERATING ROOM
To determine the patient-monitor type running on a monitor, select
MORE MENUS
>
MONITOR SETUP
>
SOFTWARE CONFIGURATION
. The patient-monitor type
is listed under
PATIENT MONITOR TYPE
.
For more information on the monitor default settings for each patient-monitor type,
refer to
Language
The monitor language is set to one of 14 options when installed. In most cases, the
language does not affect the monitor functionality or usability. However, in a few
languages, you may experience the following conditions:
NOTE
Special characters that are sent over the network in alarm broadcasts, patient
status or system status messages, patient, bed or care unit names cannot be
displayed by networked equipment that does not support these languages.
HUNGARIAN
,
POLISH
and
RUSSIAN
— Special characters may appear as
blanks, asterisks or square boxes due to the inability to display Cyrillic (Russian)
characters, as well as certain Polish and Hungarian characters.
CHINESE
and
JAPANESE
— All text for alarm broadcast sent over the
network is in English. All text input (for example, unit name, bed name, patient
information, etc.) is also in English.
CHINESE
— Invasive and non-invasive pressure parameter values are displayed
as mmHg. A converted kilopascals (kPa) value also displays in the invasive and
non-invasive pressure parameter windows when kPa is turned on. To set the
default display of the kPa values to
ON
or
OFF
, select the
MONITOR
DEFAULTS
>
SETUP DEFAULT DISPLAY
>
NBP/IBP KPA UNITS
setting.
To individually turn on or turn off the display of kPa units, select the
KPA
UNITS
control setting from the pressure parameter menu. Choosing
OFF
displays the parameter value in mmHg.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......