Monitoring ECG
2000966-386D
Dash™ 3000/4000/5000
8-5
After assessing the information extracted from each of the physiologic signals, the
IntelliRate algorithm applies rule-based logic to determine which heart rate source
has the highest likelihood of being accurate. By reporting the most accurate rate, the
trended heart rate is more accurate, and occurrences of false heart rate limit violation
alarms are greatly reduced.
The IntelliRate value replaces the standard heart rate value in the
ECG
parameter
window. It also indicates the source of the heart rate (
SPO
2
or
ART
) when the heart
rate source is not
ECG
.
To determine if IntelliRate is enabled on this monitor, select
MORE MENUS
>
MONITOR SETUP
>
SOFTWARE CONFIGURATION
. If it is enabled,
ECG
INTELLIRATE
is listed under the
ENABLED SOFTWARE OPTIONS
list.
Combo and Rover Combo Monitoring Guidelines
The following are guidelines to remember when monitoring in
COMBO
or
ROVER
COMBO
monitoring modes.
When monitoring ECG from telemetry:
ECG limits and arrhythmia alarm levels are not your monitor defaults, but
are the telemetry defaults from the central station. You can adjust these
settings at the monitor.
The alarm pause feature (if available on your transmitter) is honored at the
monitor.
You should not turn off the monitor until you have discharged the patient
from the monitor.
Second V lead data is not sent to the monitor from the telemetry transmitter.
If you wish to see telemetry second V lead data, you must view the telemetry
patient.
When switching ECG monitoring from the monitor to telemetry:
Arrhythmia alarm histories from the monitor are merged in the telemetry
system. CD Telemetry-LAN software version 5 or later is required.
If you discharge the monitor, the telemetry arrhythmia alarm levels will be
the same as the arrhythmia alarm levels supported by the monitor.
Therefore, when the monitor uses the BASIC software package, only lethal
arrhythmia alarm levels will be detected from telemetry. If the monitor has
the
CARDIAC
software package, full arrhythmia alarm levels will be
detected from telemetry.
When switching ECG monitoring from telemetry to the monitor:
Telemetry is automatically discharged and the most recent 36 alarm histories
are transferred to the monitor.
The ECG limits, arrhythmia alarm levels and display defaults are recalled
from the monitor defaults.
NOTE
It is not likely that the
COMBO
or
ROVER
COMBO
monitoring modes are used
when the patient-monitor type is
OPERATING ROOM
.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......