![GE Dash 3000 Operator'S Manual Download Page 132](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578132.webp)
10-8
Dash™ 3000/4000/5000
2000966-386D
Monitoring NBP
CUFF SIZE
Select the type of NBP cuff used on the patient (
ADULT
,
PEDIATRIC
or
NEONATAL
).
WARNING
—The correct cuff size must be selected to obtain reliable
NBP data and to prevent overpressure in neonatal or
pediatric use.
NOTE
You must select
CLEAR NBP READINGS
after
changing this setting for the target pressure to take
affect.
CLEAR NBP READINGS
Remove all values from the
NBP
parameter window, remove
NBP messages from the display and remove all NBP
information from the
VITAL SIGNS
history.
SETUP CUSTOM AUTO
CAUTION
—Changes made to these control settings are persistent.
They are recalled when a patient is discharged from the
monitor.
NOTE
Only available in software version 6 or later.
Create a custom series of automatic NBP measurements,
including the time interval between measurements and the
number of measurements to perform. For more information,
refer to
Setup Custom Automatic NBP Measurements on
INITIAL INFLATION
PRESSURE
Select the target inflation pressure for the first NBP
measurement. The choices vary by cuff size.
NOTE
You must select
CLEAR NBP READINGS
after
changing this setting for the target pressure to take
affect.
NOTE
SuperSTAT NBP briefly over-inflates beyond the initial
inflation pressure and then steps down to the correct
pressure. To benefit patient comfort, lower the default
initial inflation pressure.
PULSE RATE
NOTE
Only available in software version 6 or later.
Turn the feature on or off to display pulse rate values in the
parameter window when the display setting is
INDV 3 WFS
.
Option
Function
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......