![GE Dash 3000 Operator'S Manual Download Page 178](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578178.webp)
16-8
Dash™ 3000/4000/5000
2000966-386D
Monitoring Anesthesia Gases
2.
In
SETUP DISPLAY
locate
DISABLE SAM HAL
and
DISABLE SAM ENF
and select
ON
.
Troubleshooting
For clinical-level troubleshooting, refer to the GE “Critical Care Monitoring Clinical
Reference and Troubleshooting Guide.”
Message
Cause
Solution
BLOCKED LINE
The sample or exhaust line blocked.
Clean or replace sample or exhaust line.
CALIBRATE
The SAM module is not calibrated.
Calibrate the module.
MODULE CELL SERVICE CO2
The SAM module failed.
Contact service.
CP BYPASS
The alarms are deactivated for bypass.
This is a normal procedure message.
DESFLURANE?
Desuflurane is selected, but isoflurane is
detected.
Change the gas agent selection.
ISOFLURANE?
Isoflurane is selected, but desoflurane is
detected.
CONNECT AQUA KNOT
The Aqua-Knot water trap is disconnected.
Connect or replace the Aqua-Knot water
trap.
LOW TEMP
Gas cells are liquefied.
Apply power and wait for the module to
warm up (30 minutes to one hour).
GAS LIQUEFIED/WARMING UP
Due to extreme temperature change, gas in
one or more of the internal cells is liquefied.
NO BREATH DETECTED
No breath detected or sample line
disconnected.
Check the patient and the sensor.
AIR CAL IN PROGRESS
Room air calibration in progress.
This is a normal procedure message,
SERVICE…
Unknown hardware failure.
Contact service.
UNKNOWN ERROR
WARMING UP
The SAM module is warming up.
If the message displays for more than four
minutes, replace module or contact service.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......