Monitoring Respiration
Troubleshooting
For clinical-level troubleshooting, refer to the GE “Critical Care Monitoring Clinical
Reference and Troubleshooting Guide.”
Option
Function
LEAD
Select the lead to derive the respiration rate.
RELEARN RESPIRATION
Relearn the patient’s respiration pattern over the next eight
breaths. While in process, the message “
LEARNING
”
displays. When complete, the respiration rate value displays
at 40 percent sensitivity and the waveform is automatically
sized.
SENSITIVITY
Select the detection sensitivity of the average amplitude.
The lower the value, the greater the detection sensitivity.
RESPIRATION LIMITS
Select the high and low alarm levels for respiration rate and
apnea. An alarm sounds when a value occurs outside of
these limits.
AUTO SIZE
Select the size of the displayed waveform to fit as factory-
intended on the display. The waveform size has no effect on
detection.
MANUAL SIZE
Select the size of the displayed waveform. The waveform
size has no effect on detection.
CARDIAC ARTIFACT
ALARM
Turn this patient status
ADVISORY
alarm on or off. It alerts
you when the respiration rate is within five percent of the
heart rate over 30 consecutive breaths because the monitor
may be counting heart beat artifact as respiration. It displays
the message “
ARTIFACT
” in the
RR
parameter window and
sounds a one beep tone.
SPEED
Select a sweep speed for the displayed waveform.
Message
Cause
Solution
LEARNING
The monitor is relearning the patient’s
respiration pattern.
This is a normal procedure message.
APNEA
No breath detected.
Check the patient and the sensor.
ARTIFACT
The monitor cannot determine respiration.
Check the patient and sensor.
LD I (II) FAIL
The monitor cannot detect patient data from
the current lead.
Check the patient, and replace or switch the
sensor.
LEADS FAIL
The monitor cannot detect patient data from
any lead.
Check the patient, and replace or switch the
sensor.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......