![GE Dash 3000 Operator'S Manual Download Page 175](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578175.webp)
Monitoring Anesthesia Gases
2000966-386D
Dash™ 3000/4000/5000
16-5
NOTE
Always position the Luer-Lok connection and patient sample line in an upright
position to avoid aspiration of fluids into the patient sample line, which will
cause premature blockage of the Aqua-Knot water trap.
NOTE
Delivery of bronchodilators, mucolytics via aerosol, and meter dose inhalers will
cause premature blockage of the Aqua-Knot water trap. Discontinue gas analysis
prior to treatment by removing the airway adapter from the ventilator circuit, or
by disconnecting the sample line from the Aqua-Knot water trap.
Monitoring
To verify that you are ready to monitor anesthetic gases, follow this procedure.
1.
Install the SAM module into the TRAM-RAC 2A module housing. You will hear
a click when the module is fully inserted.
2.
Connect one end of the TRAM-NET communication cable into the
Auto
or
TRAM-NET
port on the back of the module housing.
3.
Connect the other end of the communication cable into the
Aux
port on the back
of the monitor.
4.
Turn on the power switch on the back of the module housing.
5.
Turn on the monitor.
6.
Admit the patient to the monitor.
7.
Position the sensor on the patient.
8.
Verify that patient data displays in the
CO
2
and
GAS
parameter windows.
9.
Adjust the control settings as necessary.
Control Settings
CO
2
To adjust the control settings for the patient admitted to the monitor, select the
CO
2
parameter window label. Then select any of the following options and make the
desired changes.
NOTE
All changes are temporary and revert to the factory or custom defaults when the
patient is discharged. To permanently change the these settings, refer to
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......