Monitoring NBP
2000966-386D
Dash™ 3000/4000/5000
10-7
NOTE
All changes are temporary and revert to the factory or custom defaults when the
patient is discharged. To permanently change the these settings, refer to
Option
Function
NBP AUTO
Turn automatic determination off or select the time interval to
take automatic NBP measurements. The default setting is
five minutes.
A countdown timer displays in the NBP parameter window
when the time interval is less than 60 minutes.
To synchronize NBP measurements with actual clock times,
change the
MONITOR DEFAULTS
>
SETUP DEFAULT
DISPLAY
>
NBP CLOCK SYNC
option to
ON
.
WARNING
—Periodically check patient limb circulation distal to the cuff.
Check frequently when using Auto NBP in one or two minute
intervals. Intervals below 10 minutes are not recommended
for extended periods of time.
CAUTION
NBP AUTO MODE SHUT OFF—When you select the
GO
TO STANDBY
option or press the
Standby
key, the
NBP
AUTO
mode automatically shuts off.
NBP STAT
NOTE
Not available when in the
NEONATAL-ICU
patient-
monitor type.
Start five minutes of automatic, continuous, sequential NBP
measurements.
When initiated, cuff inflation begins, a measurement is taken
and the systolic, diastolic and mean pressure values are
displayed. Thereafter, the systolic value displays after the
measurement starts, but before the other pressure values
are complete. The monitor beeps when all values are
complete.
Press the
NBP Go/Stop
key to stop the measurements.
REVIEW NBPS
Review the previous 96 NBP measurements less than 24
hours old in the
VITAL SIGNS
information window.
NBP LIMITS
Select the high and low alarm limits for systolic, diastolic and
mean pressures. An alarm sounds when a value occurs
outside of these limits.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......