![GE Dash 3000 Operator'S Manual Download Page 121](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578121.webp)
Monitoring Invasive Pressures
2000966-386D
Dash™ 3000/4000/5000
9-5
Option
Function
SCALES
Select the size of displayed waveforms. The higher the scale
value, the smaller the waveform scale.
AUTO
— Adjusts the scale to patient’s ART pressure,
Auto scaled waveforms print at a slightly different scale
than the display scale.
FULL
— Allows you to select the scale size and
position on the display; right (
R
) or left (
L
).
CURSOR
Display a horizontal line with a numeric value over the
waveform to indicate a pressure value at specific position on
the waveform.
CLEAR CURSOR
Remove the cursor from the waveform.
LIMITS
Select the high and low alarm limits for heart rate, systolic,
diastolic and mean pressures. An alarm sounds when a
value occurs outside of these limits.
NOTE
Adjusting the HR limits affects the pressure tracing, not
the primary ECG HR alarm level.
CHANGE NAME
Select the site name for the monitored pressure site for
accurate waveform processing (for example, if monitoring
the femoral pressure site, select
FEM
).
ZERO
Zero a single transducer. To zero all transducers, press the
Zero All
key.
BP FILTER
Select one of the following filter options:
12 HZ
— Recommended for typical monitoring.
40 HZ
— Recommended for higher frequency
waveform component processing. May result is
elevated pressure values.
CALIBRATE
TRANSDUCER
Adjust the monitor to compensate for transducer errors.
Measurement errors can be identified when pressure
calibration is measured with other instruments (for example,
a manometer).
CAL# 100
— Select the percentage of error when the
calibration error is known; (for example, 102 if the
transducer measured two percent high).
ADJUST TO MMHG
— Select the millimeters of
mercury.
SPEED
Select a sweep speed for the displayed waveform. The
smaller the speed value, the slower the speed (greater
number of complexes).
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......