Monitoring Bispectral Index (BIS)
2000966-386D
Dash™ 3000/4000/5000
18-5
NOTE
All changes are temporary and revert to the factory or custom defaults when the
patient is discharged. To permanently change the these settings, refer to
Disabling Continuous Lead Detection
By default, the BISx continuously checks electrode impedances. This 1 nA 128 Hz
impedance check signal may interfere with other equipment, such as evoked
Option
Function
SMOOTHING
Select the averaging time used to calculate BIS: The factory
default setting is 15 seconds.
15 SEC
— Select for operating room or sedation
patient monitoring to increase the responsiveness to
changes in the patient’s state.
30 SEC
— Select for intensive care patient monitoring
to provide a smoother BIS trend with less variability
and sensitivity to artifact.
BIS LIMITS
Select the high and low alarm limits for BIS parameter. An
alarm sounds when a value occurs outside of these limits.
SENSOR CHECK
Select to automatically detect sensors and display the
impedance value and status of each lead. This data is
updated every two seconds.
PASS
— This lead passed the sensor check.
LD OFF
— No impedance value; check the lead.
NOISE
— No impedance value; check the lead.
HIGH
— Re-prep the sensor.
BIS SETUP
Provides access to the following controls:
LEAD DETECT
— For more information, refer to
Disabling Continuous Lead Detection on page 18-5
.
EEG FILTER
— For more information, refer to
Disabling the Waveform Filter on page 18-6
.
BIS TEST
— For more information, refer to
.
RESUME BIS
Select to initiate a sensor check when the “
EXPIRED
SENSOR
” message displays.
EEG WAVEFORM
Select
ON
to display EEG waveform data.
EEG SCALE
Select the scale of the displayed waveform. The larger the
scale, the smaller the waveform. The factory default setting
is
25 µV/DIV
.
BIS RANGES
Displays a description of the factory default BIS ranges.
BIS TRENDS
Display the trended BIS values in the
VITAL SIGNS
information window.
SPEED
Select a sweep speed for the displayed waveform.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......