5-12
Dash™ 3000/4000/5000
2000966-386D
Managing Patients
The host monitor is completely functional and two seconds of waveforms data
continue to display. The menus still control the host monitor and any changes you
make do not affect the function of the viewed monitor. Information windows will also
display over the viewed monitor.
Keep in mind the following guidelines when viewing a split view:
Admit a patient to view. If the monitor is discharged, the parameter windows
display (no waveforms) with a “
DISCHARGED
” message.
To maintain CO
2
monitoring on the host monitor while viewing another monitor,
change the position of CO
2
to the sixth position. The slower CO
2
sweep speed
will block all other waveforms adjacent or below the CO
2
position.
CO
2
waveform and parameter windows are not sent to the host monitor.
ST trends and complexes are not sent to the host monitor.
CRG+ display is not sent to the host monitor.
Removing a Viewed Monitor Display
To remove an alarming monitor display from the host monitor:
VOA — Select
MAIN MENU
or
RETURN
.
AVOA — Perform any one of the following actions:
Select
MAIN MENU
or
RETURN
.
Silence the alarming monitor at the host monitor or central station.
Resolve the alarm condition at the alarming monitor (all alarm levels except
CRISIS
).
Silence the remote alarm. This also prevents new alarms from sounding,
even if a new alarm is of higher priority.
AutoView XM — Perform any one of the following actions:
Select
MAIN MENU
or
RETURN
.
Silence the alarming monitor at the host monitor or central station.
Resolve the alarm condition at the alarming monitor (all alarm levels except
CRISIS
).
Silence the remote alarm when the host monitor’s
AVOA SILENCE
setting
is set to
SILENCED
. This also prevents new alarms from sounding, even if
a new alarm is of higher priority.
A more recent, higher priority alarm is received at the host monitor when the
host monitor’s
AVOA SILENCE
setting is set to
NEW ALARM
.
NOTE
To remove a non-alarming monitor display from the host monitor, select
MORE
MENUS
>
VIEW OTHER PATIENTS
>
SELECT A BED TO VIEW.
The
viewed bed is indicated with the word “
VIEW”
. Selecting that bed again will
remove from the host monitor.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......