Locale
The monitor is set to one of two locale options when installed:
DEFAULT
or France.
If the monitor is set to
FRANCE
, the following functionality varies from the
standard, default monitor functions.
France
Alarms
Alarm pause — Lasts no more than three minutes when in
ADULT-ICU
or
NEONATAL-ICU
patient-monitor types.
Alarm volume — Alarms can not be turned off in
NEONATAL-ICU
patient-
monitor type. The lowest possible setting is 10 percent.
When turned off, the monitor alarms every three minutes to remind you alarms
are silenced. The volume resets to the monitor default setting when a patient is
admitted, alarm pause breakthrough occurs or the monitor is turned on.
Discharge alarms — The monitor alarms every three minutes when a patient is
discharged.
Alarm levels — All parameter patient status message level alarms (except ECG)
are automatically increased to a patient status advisory level alarm.
The following monitor default alarm limits are also unique:
NBP-S, NBP-D, NBP-M
(
ADULT-ICU
and
OPERATING ROOM
) — 350 to -
25.
NBP-S, NBP-D, NBP-M
(
NEONATAL-ICU
) — 150 to -25.
ART
— 300 to 30.
HR
— 300 to 30.
SPO2
— 105 to 60.
SPO2-R
— 235 to 40.
TEMP
— 450 to 0.
Parameters
The following parameter values are unique:
ECG gain — Derived by mV (.25 mV to 2 mV) and displayed on the primary
ECG waveform.
Units of measurement — Metric only.
Arrhythmia level — The default arrhythmia alarm level for
ADULT-ICU
is
FULL
;
NEONATAL-ICU
and
OPERATING ROOM
are
OFF
.
Heart rate — Values under 30 bpm are not displayed. The values displayed in the
parameter window are replaced with “
X
”.
SpO
2
saturation — Values under 60 percent are not displayed. The values
displayed in the parameter window are replaced with “
X
”.
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......