![GE Dash 3000 Operator'S Manual Download Page 249](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578249.webp)
Interfacing with Non-GE Devices
2000966-386D
Dash™ 3000/4000/5000
E-13
One transcutaneous monitor.
One urometer.
One continuous cardiac output monitor.
One multi-parameter monitor.
Two pulse oximeters.
One auto-association device (GE monitor).
Duplicate Parameters
The connectivity device can support a pre-defined number of parameters. When the
maximum number is exceeded, the connectivity device ignores all data from the
duplicate parameter.
Two SpO
2
parameters.
Four temperature parameters.
Eight BP parameters.
Eight IV parameters.
One CO
2
or Gas parameter.
One each of all other parameters
CO
2
and Gas Parameter Handling
The connectivity device requires that CO
2
and Gas parameters originate from the
same peripheral device. If a device only provides one of the parameters, the
connectivity device does not recognize the other parameter from any other peripheral
device.
However, when the connectivity device is connected to a Dash 3000/4000/5000
monitor, the monitor controls which source of CO
2
and Gas parameter data is used.
The monitor does NOT require that CO
2
and Gas parameters originate from the same
peripheral device.
Refer to the following table for more information CO
2
and Gas parameter data.
SAM Module
Monitor CO
2
Option
Unity Network ID
CO
2
Data Source
Gas Data Source
Yes
No
No
SAM module
SAM module
Yes
Yes
No
SAM module
SAM module
Yes
No
Yes
SAM module
SAM module
Yes
Yes
Yes
SAM module
SAM module
No
Yes
No
Monitor
None
No
Yes
Yes
Monitor
Unity Network ID
No
No
Yes
Unity Network ID
Unity Network ID
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......