MM 95:
The concentrate connectors are connected to the
water flow path by self-sealing valves.
The heated water flows through the feeding
pumps. The pumps are ceramic, and need a
constant flow of water through the pump ele-
ment; this acts as a lubricant.
In Bicarbonate mode the acid concentrate is
taken up through the red A-concentrate connec-
tor by the
A-pump
and added to the water flow.
The acid/water mixture goes through a
mixing
chamber,
to prevent air from causing unstable
conductivity readings, before entering the
A-
conductivity cell
. The measurement from the
A-cell is compared with the required conductiv-
ity and any difference between the two will
result in a change of the A-pump speed to correct
the difference. There is a temperature trans-
ducer in the conductivity cell to compensate the
conductivity calculation since it varies with
temperature, approxi1,8 % per °C.
The bicarbonate concentrate is taken up through
the B-concentrate connector by the
B-pump
and
added to the acid/water solution. The acid/
bicarbonate/water mixture goes through the
second
mixing chamber
, to prevent air from
causing unstable conductivity readings, before
entering the
B- conductivity cell
. The meas-
urement from the B-cell is compared with the
required conductivity and any difference be-
tween the two will result in a change of the
B-pump speed to correct the difference. There is
a temperature transducer in the B-conductivity
cell to compensate the conductivity calculation
since it varies with temperature, approximately
+1,8 % per °C.
When
BiCart
®
is mounted, the flushing valve,
FLVA,
is used for priming of the BiCart car-
tridge. FLVA is also used during fill up of di-
luted chemical disinfection solution.
Note that in acetate mode the A conductivity cell
and A pump are not used. The acetate concen-
trate is taken up through the blue B-concentrate
connector.
The mixed fluid is then passing the (optional)
pH-sensor before leaving the MM 95.
B, Ac
A
B
T
Cond.
Cell "A"
T
Mixing Chamber
Mixing Chamber
Feeding
Pumps
MM 95
A
Cond.Cell "B"
Bi-
Cart
FLVA
pH-SENSOR
Figure 4. MM 95 - Flow diagram
Summary of Contents for AK 95 S
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care...
Page 5: ...This page is intentionally left blank...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank...
Page 247: ...AK 95 AK 95 S Service Manual...