Electrical Safety Inspections for AK-machines
11:11
5
Checklist
Electrical Safety Inspection for AK machines
- to be filed by the Service Engineer responsible
AK 95 / AK 95 S
AK 200 / AK 200 ULTRA / AK 200 S / AK 200 ULTRA S
Equipment ID:_______________Serial number:__________________
Run time (h):__________
Visual inspection
Visual inspection of the AK-machine according to section 2.
PET for AK 95, AK 95 S
(1) Resistance between the grounding clips of the mains plug and the exterior
cover of the power supply is < 0.3 Ohm.
(2) Resistance between the grounding clips of the mains plug and the heater
rod is < 0.3 Ohm.
(3) Resistance between the grounding clips of the mains plug and the suction
pump is < 0.3 Ohm.
(4) Resistance between the grounding clips of the mains plug and the mount-
ing plate to the diascan conductivity cell is < 0.3 Ohm.
(5) Resistance between the grounding clips of the mains plug and the stand
of MM 95 (e.g. a screw to the heat exchangers) is < 0.3 Ohm.
(6) Resistance between the grounding clips of the mains plug and the front
of MM 95 (e.g. a screw to the BiCart-holder)
(7) Resistance between the grounding clips of the mains plug and the heparin
pump (rear plate) is < 0.3 Ohm.
(8) Resistance between the grounding clips of the mains plug and the poten-
tial equalization connector is < 0.3 Ohm.
PET for AK 200, AK 200 ULTRA, AK 200 S, AK 200 ULTRA S
(1) Resistance between the grounding clips of the mains plug and the power
supply (right side, through the ventholes of the bottom box) is < 0.3 Ohm.
(2) Resistance between the grounding clips of the mains plug and the FM
bottom plate is < 0.3 Ohm.
(3) Resistance between the grounding clips of the mains plug and the heater
rods outside the heating vessel is <0.3 Ohm.
(4) For 115 V machines: Resistance between the grounding clips of the
mains plug for the heater rods and the heater rods outside the heating vessel
is < 0.3 Ohm.
(5) Resistance between the grounding clips of the mains plug and the over-
heat protection of the heater rods is < 0.3 Ohm.
(6) Resistance between the grounding clips of the mains plug and the poten-
tial equalization connector is < 0.3 Ohm.
(7) Resistance between the grounding clips of the mains plug and the heparin
pump (rear plate) is < 0.3 Ohm.
(8) Resistance between the grounding clips of the mains plug and the rear
plate of the BVS-unit is < 0.3 Ohm.
Summary of Contents for AK 95 S
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care...
Page 5: ...This page is intentionally left blank...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank...
Page 247: ...AK 95 AK 95 S Service Manual...