Technical Description
AK 95 / AK 95 S Service Manual
HCEN 9134 Rev. 10.2005
2:18
Treatment
Arterial blood is taken in via the
arterial line
clamp
. During treatment the line clamp is open.
The
arterial pressure guard
is used to give an
alarm if the pressure becomes too low. For
example, if the needle or the arterial blood line
is blocked, an alarm will be given ( the blood
pump is stopped). The alarm is activated if there
is an increased negative pressure between the
patient and the blood pump,e.g:
•
A drop in blood pressure.
•
Altered positions of the arterial needle.
•
A kink in the arterial line between the
patient and the blood pump.
The function of the
blood pump
is to maintain
the extra corporeal blood flow. Blood is removed
from the patient, forwarded to the dialyzer and
then returned to the patient.
The blood pump can not be run when the pump
cover is open.
If necessary, the blood can be returned to the
patient manually.
The purpose of the
heparin pump
is to add
heparin to the blood.
An alarm is issued when the syringe is empty or
if the heparin line is kinked.
The heparin parameters can be adjusted by
pressing the
heparin
button.
The
venous pressure transducer
measures
the venous bloodpressure. The lamp in the
Venous pressure button is lit and the buzzer
sounds if the alarms are activated and the
venous bloodpressure is outside the alarm
limits. If the event of an alarm, the blood flow is
stopped.
If the venous pressure is too high, the cause may
be:
•
An obstruction after the venous drip
chamber.
•
A change in the patients position.
If the venous pressure is too low, the cause may
be:
•
A line separation.
•
A drop in blood pressure.
•
A change in the patients position.
The
air detector
has the following function:
If the blood level in the
venous drip chamber
is too low, an alarm is given and the blood flow is
stopped (the blood pump and heparin pump are
stopped and the venous line clamp is closed). The
venous drip chamber acts as an expansion cham-
ber to even out pressure pulsations.
When the air detection alarm is generated, the
lamp in the Air detect button is lit and the
buzzer sounds. The lamp begins flashing when
the air detector does not detect air any longer.
The operator must reset the air detection alarm
by pressing the
Air detect
button. After press-
ing the
Air detect
button for 3 seconds, the air
detector alarm is bypassed for x seconds (preset-
table 0-120 sec.). Default it is disabled (0 sec-
onds) and no override is possible.
The
priming detector
detects if there is blood
in the venous blood line. If no blood is detected,
it is not possible to start the treatment.
The blood is transported back to the patient via
the
venous line clamp
. The clamp is normally
open. But in the event of an alarm from the air
detector or the venous pressure transducer, the
line clamp clamps the blood line and stops the
blood flow (not during high pressure alarm).
Summary of Contents for AK 95 S
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care...
Page 5: ...This page is intentionally left blank...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank...
Page 247: ...AK 95 AK 95 S Service Manual...